Plastic Rockwell Hardness Tester
Plastics Rockwell hardness tester adopts the mechanism of automatic loading and unloading of the test force, the test force is regulated by the load-change hand wheel. It can select the dwell time and is very easy to use. Besides setting zero to dial, there is no man-made error. The instrument has high sensitivity and stability.
Main Application:
Suitable for hard plastics, hard rubbers, aluminum, tin, copper, soft steel, synthetic resins and fricative
materials etc.
EI-R150EP: Electric Dial Type
- The dial will display hardness value directly,
indicator responses sensitively, show the hardness value accurately. - Cast molding frame, stable and robust structure.
- Preliminary test force load by manual, main test force load by motor.
- Automatic Load – Dwell – Unload main test force
- Dwell time can be pre-setting (2 ~ 60s)
EI-R150TP: Digital Touch Screen
- Electric loading with weights, high precision sensor.
- Touch screen interface, easy operation.
- Automatic correction of high, medium and low hardness values.
- Built-in hardness value correction of convex cylindrical surface.
- Strong data processing function, conversion of hardness value HR, HB, HV and so on, set the qualified range, overrun alarm, sample name and other information.
- Test results can be stored, processed and printed automatically.
Specification
Model | EI-R150EP | EI-R150TP |
Hardness Scale | HRE, HRL, HRM, HRR | HRA, HRB, HRC, HRD, HRE, HRF, HRG, HRH, HRK, HRL, HRM, HRR,HRP, HRS, HRV |
Preliminary Test Force | 10kg (98.07N) | |
Test Force | 60kg (588.4N), 100kg (980.7N), 150kg (1471N) | |
Conversion Scale | – | HRA, HRB, HRC, HRD, HRE, HRF, HV, HBW, HR15N, HR30N, HR45N, HR15T, HR30T, HR45T |
Dwell Time | 2 ~ 60 seconds | 0 ~ 99 seconds |
Resolution | 0.5 | 0.1 |
Accuracy | ± 1 % | < 0.5 % |
Hardness Reading | Analog Dial | 8” LCD Touch Screen Display |
Output | – |
|
Test Space | Max. Height: 175mm Max. Throat Depth:165mm |
Max. Height: 200mm Max. Throat Depth:160mm |
Indenter | Ø3.175mm, Ø6.35mm, Ø12.7mm Ball Indenter |
Diamond Rockwell Indenter & 1.5875mm Ball Indenter |
Test Force Change | By change weights | |
Dimension | 520 x 215 x 700 | 560 x 200 x 800 |
Weight | Approx. 78kg | Approx. 70kg |
Accordance standard | ISO 6508, ASTM E-18, JIS Z2245, GB/T 230.2 |